Product

12.1″ Patient Monitor S9000

12.1" color TFT screen; Multi‐channel display ; Big font interface; ST and arrhythmia analysis; Drug dose calculation and titration table; OxyCRG dynamic view display; Carton size: 42cm*26cm*37cm; NW: 3.5kgs; CE and ISO

12.1" color TFT screen

Multi‐channel display

Big font interface

ST and arrhythmia analysis

Drug dose calculation and titration table

OxyCRG dynamic view display

Respiration monitoring with apnea alarms

Alarm review, arrhythmia review

Patient info. Input

Anti‐defibrillation design

NIBP over pressure protection

Plug‐in rechargeable battery, AC/DC available.

Supporting wire or wireless network.


Technical Specifications

ECG

Lead Mode                          5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)

Lead selection                     I, II, III, avR, avL, avF, V

Wavef orm                          2 ch

Lead mode                          3 Leads ( R, L, F or RA, LA, LL)

Lead selection                     I, II, III,

Wave form                          1 ch

Gain                               2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto

 

HR and Alarm Range                 Adult    15 ~ 300 bpm

                                   Neo/Ped    15 ~ 350 bpm

Accuracy                           ±1% or±1bpm, which is greater

Resolution                         1 bpm

Sensitivity                        > 200 (uVp-p)

Differential Input Impedance       > 5 MΩ

CMRR                   Monitor > 105 dB

                                   Operation > 105 dB

                                   Diagnosis > 85 dB

Electrode offset potential         ±300mV

Leakage Current                    < 10 uA

Baseline Recovery                  < 3 S After Defi.

ECG Signal Range                   ±8 m V ( Vp-p )

Bandwidth                           Surgery    1 ~ 20 Hz

                                    Monitor     0.5 ~ 40 Hz

                                    Diagnostic  0.05 ~ 130 Hz

Calibration Signal                  1 (mVp-p),  Accuracy:5%

ST Segment Monitoring Range         Measure and Alarm -2.0 ~ +2.0 mV

ARR Detecting                       Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,

                                          PNP, PNC,Alarm Available,Review Available

Respiration

Method                            Impedance between R-F(RA-LL)

Differential Input Impedance       > 2.5 MΩ

Measuring Impedance Range          0.3~3 Ω

Base line Impedance Range           200~4 KΩ

Bandwidth                          0.1~2.5Hz

Resp. Rate                         Measuring and Alarm Range

                                    Adult 0 ~ 120 Brpm

                                    Neo/Ped 0 ~ 150 Brpm

                                    Resolution 1 Brpm

                                    Accuracy ±2 Brpm

Apnea Alarm                        10 ~ 40 S

NIBP

Method                            Oscillometric

Mode                              Manual, Auto, STAT                        

Measuring Interval in AUTO Mode   1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)

Measuring Period in STAT Mode     5 Min

Pulse Rate Range           40 ~ 240 bpm

Alarm Type                  SYS, DIA, MEAN

Measuring and alarm range        Adult Mode

                                    SYS 40 ~ 270 mmHg

                                    DIA 10 ~ 215 mmHg

                                    MEAN 20 ~ 235 mmHg

                                Pediatric Mode

                                    SYS 40 ~ 200 mmHg

                                    DIA 10 ~ 150 mmHg

                                    MEAN 20 ~ 165 mmHg

                                Neonatal Mode

                                    SYS 40 ~ 135 mmHg

                                    DIA 10 ~ 100 mmHg

                                    MEAN 20 ~ 110 mmHg

Resolution                      1mmHg

Accuracy                        Maximum Mean error ±5mmHg

                                Maximum Standard deviation ±8mmHg

Overpressure Protection          Adult Mode 297±3 mmHg

                                 Pediatric Mode 240±3 mmHg

                                 Neonatal Mode 147±3 mmHg

SpO2

Measuring Range                  0 ~ 100 %

Alarm Range                      0 ~ 100 %

Resolution                       1 %

Accuracy                         70% ~ 100%  2 %

                                 0%  ~  69% unspecified

Actualization interval about      1 Sec.

Alarm Delay                      10 Sec.

Pulse Rate                       Measuring and Alarm Range 20~300 bpm

                                 Resolution       1bpm

                                 Accuracy ±3bpm                                 

TEMP

Channel                          2

Measuring and Alarm Range 0 ~ 50 ℃

Resolution                      0.1 ℃

Accuracy                 ±0.2 ℃

Actualization intervalabout     1 Sec.

Average Time Constant           < 10 Sec.

IBP

Label                           ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measuring and alarm range       ART 0 ~ 300 mmHg

                                PA -6 ~ 120 mmHg

                                CVP/RAP/LAP/ICP-10 ~ 40 mmHg

                                P1 / P2-10 ~ 300 mmHg

Press Sensor                    Sensitivity 5  uV/V/mmHg

                                Impedance 300-3000Ω

Resolution                      1 mmHg

Accuracy                        ±2% or 1mmHg which is greater

Actualization interval         about 1 secretary

 

Standard:  ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature

Optional:  Nellcor SpO2, EtCO2, IBP,TEMP-2, Thermal Recorder, Wall Mount,Trolly, Central station