Product

8″ Patient Monitor S6000B

8" color TFT LCD screen, resolution 800*600; Waterproof design, resistance to dropping (1 meter) ; Low power consumption, fanless structure which creates no noise;15 kinds of arrhythmia analysis, 15 kinds of drug concentration analysis; Carton size: 40cm*15cm*28cm; NW: 2.6; CE and ISO

●  8" color TFT LCD screen, resolution 800*600

●  Waterproof design, resistance to dropping (1 meter)

●  Low power consumption, fanless structure which

     creates no noise

● Support external 13-20V DC power, continuously working

     for 5.5 hours above on battery backup, full charge with 2.5

     hours, suitable for Ambulance, Emergency Transport

●  Display mode: Big font, trends review, 7 leads ECG waveform

      on same screen, and etc

●  15 kinds of arrhythmia analysis, 15 kinds of drug concentration

      analysis

●  Anti-electrosurgical unit (300W), anti-defibrillation (360J),

      suitable for Operating Room, General Ward

●  Digital SpO2 technology, excellent anti- interference

      performance, low perfusion (0.1%)

ECG

Lead: Standard 3/5 lead

Lead Mode: 5 Leads (RA, LA, LL, RL,V)

Lead Selection: I, II, III, avR, avL, avF, V

Lead Mode: 3 Leads (RA, LA, LL)

Lead Selection: I, II, III,

Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV

HR and Alarm Range:

        Adult: 15 ~ 300 bpm

        Ped/Neo: 15 ~ 350 bpm

Accuracy: ±1% or±1bpm, which is greater

Resolution: 1 bpm

Sensitivity: > 200 (uVp-p)

Differential Input Impedance: > 5 MΩ

Bandwidth:

        Surgery    1 ~ 20 Hz

        Monitor     0.5 ~ 40 Hz

        Diagnostic  0.05 ~ 130 Hz

CMRR:

        Monitor > 100 dB

        Operation > 100 dB

        Diagnosis > 90 dB

Electrode Offset Potential: ±300mV

Range: ±2 mV ~ ±700 mV

Width: 0.1 ms ~ 2 ms

Rise time: 10 us ~ 100 us

ST Segment Monitoring Range:

        Measurement Range: -2.0 ~ +2.0 mV

        Measurement Accuracy: -0.8 ~ +0.8 mV

        Measurement Error: ±0.02 mV or ±10% 

RESP

Method: Impedance Between RA-LL or RA-LA

Measuring Impedance Range: 0.3~3 Ω

Base line Impedance Range: 200~4 KΩ

Bandwidth: 0.1~2.5Hz

Resp. Rate:

        Measuring and Alarm Range:

            Adult: 2 ~ 120 BrPM

            Neo/Ped: 2 ~ 150 BrPM

            Resolution: 1 BrPM

            Accuracy: ±2 BrPM

            Apnea Alarm: 10 ~ 40 S 

NIBP

Method: Oscillometric

Mode: Manual, Auto, STAT                        

Measuring Interval in Auto Mode:

        1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)

Measuring Period in STAT Mode: 5 Min

Pulse Rate Range: 40 ~ 240 bpm

Measuring and Alarm Range:

        Adult Mode

            SYS 40 ~ 270 mmHg

            DIA 10 ~ 215 mmHg

            MEAN 20 ~ 235 mmHg                       

        Pediatric Mode

            SYS 40 ~ 200 mmHg

            DIA 10 ~ 150 mmHg

            MEAN 20 ~ 165 mmHg

        Neonatal Mode

            SYS 40 ~ 135 mmHg

            DIA 10 ~ 100 mmHg

            MEAN 20 ~ 110 mmHg

Static Pressure Range: 0 ~ 300 mmHg

Static Pressure Accuracy: ±3mmHg

Accuracy: Maximum Average Error ±5mmHg

                  Maximum Standard Deviation ±8mmHg

Overpressure Protection:

        Adult Mode: 300 mmHg

        Pediatric Mode: 240 mmHg

        Neonatal Mode: 150 mmHg

SpO2

Measuring Range: 0 ~ 100 %

Resolution: 1 %

Accuracy: 70% ~ 100%  ±2 %

                  0% ~ 69% unspecified

Pulse Rate:

        Measuring Range: 20~300 bpm

        Resolution: 1bpm

        Accuracy: ±3bpm

TEMP

Applicable Temperature Sensor: YSI, CYF

Channel: 1

Measuring Range: 0 ~ 50 ℃

Resolution: ±0.1 ℃

Accuracy: ±0.1 ℃ (Do not including sensor error)

Dimension: 210*80*175mm (L*W*H)

Weight: 1.35kg

 

Standard: 3/5-Lead ECG, RESP, SpO2, NIBP, PR, 1-TEMP

Optional: Mainstream/Sidestream EtCO2, Touch screen

Click to Contact